I-Tech T-ONE REHAB Notice D'utilisation page 107

Masquer les pouces Voir aussi pour T-ONE REHAB:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ATTENTION. Do not use the iontophoresis program in proximity of
metal prostheses.
REHA1-2-3 • Ionophoresis L-M-H (medical program)
The patient may experience a slight reddening of the skin at the end of the
program; the redness should disappear a few minutes after the end of the
program.
Channel 2 is disconnected.
Session duration: 30 minutes.
Electrodes' positioning: place the electrode with the medicine on the painful
area and the other one transversely.
Intensity: for the iontophoresis programs the intensity must be adjusted so
that the patient experiences a significant tingling sensation in the treated
area, producing a slight contraction of the muscles surrounding it.
REHA 4 • Micro-current (medical program)
The use of the micro-current is similar to the conventional TENS, with the only
difference that sometimes, the electrical current can be more easily adapted
to the sensitivity of patients suffering from mild anxiety or to delicate body
parts.
In general, it can be used for daily pains but if the origin of the pain is
unknown, you should contact your doctor.
It can be considered as an analgesic current ideal for multiple applications,
without side effects (except for a mild redness of the skin after long
applications) and with few contraindications (those listed in the
Contraindications paragraph)
Session duration: 30 minutes.
Electrodes' positioning: over the painful area as shown in Figure 1.
Intensity: adjusted just above the threshold of perception.
REHA 5 • Haematomas (medical program)
Consult your doctor before using this program on haematomas. Few
applications within a few hours from the contusion. Combining different
types of square wave pulses, this program increases microcirculation and
drains the affected area (you can change the depth of drainage by changing
the frequency of the pulses).
Session duration: 30 minutes.
Electrodes' positioning: form a square over the area to be treated, as shown
in Figura 1.
IACER Srl
103
MNPG236-02

Publicité

Table des Matières
loading

Table des Matières