OUTPUT
Voltage Range:
Current Range:
Output Voltage Setting:
Output Current Setting:
Operating Mode:
Output Switch:
Output Terminals:
Sensing:
Output Protection:
Over-temperature
Protection:
Line Regulation:
Load Regulation:
Ripple & Noise
(20MHz bandwidth):
Transient Response:
Temperature Coefficient:
Status Indication:
METER SPECIFICATIONS
Meter Types:
Meter Resolutions:
Meter Accuracies:
GENERAL
AC Input:
Power Consumption:
Operating Range:
Storage Range:
Environmental:
Safety & EMC:
Size & Weight:
2
0V to 20V minimum
0A to 20A minimum
By coarse and fine controls.
By single logarithmic control.
Constant voltage or constant current with automatic cross-over.
Electronic. Preset voltage and current displayed when off.
Universal 4mm safety binding posts on 19mm (0.75") pitch.
Switchable between local and remote. Spring-loaded push
terminals for remote connection.
Output will withstand up to 25V forward voltage.
Reverse protection by diode clamp for reverse currents up to 3A.
The output will be tripped off if a fault causes the internal
temperature to rise excessively.
Change in output for a 10% line change:
Constant voltage: <0.01% + 2mV
Constant current: <0.01% + 2mA
Change in output for any load change, remote sense connected:
Constant voltage: <0.01% + 5mV
Constant current: <0.05% + 2mA
Typically <2mVrms, <20mV pk-pk, constant voltage mode.
<0.5ms to within 50mV of set level for a 5% to 95% load change.
Typically <100ppm/°C
Output on lamp. Constant current mode lamp.
Dual 4 digit meters with 14mm (0.56") LEDs. Reading rate 4 Hz.
10mV, 10mA
Voltage 0.3% of reading ± 3digits,
Current 0.6% of reading ± 3digits
220-240VAC ±10%, 50/60Hz; 110-120VAC ± 10%, 50/60Hz by
factory option. Installation Category II.
800VA max.
+5ºC to +40ºC, 20% to 80% RH.
40ºC to + 70ºC.
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Complies with EN61010-1 & EN61326-1.
For details, request the EU Declaration of Conformity for this
instrument via http://www.aimtti.com/support (serial no. needed).
140 x 160 x 320mm (WxHxD), 3.6kg.