General specifications apply for the temperature range 5°C to 40°C. Accuracy specifications
apply for the temperature range 18°C to 28°C after 1 hour warm-up with no load and calibration
at 23°C. Typical specifications are determined by design and are not guaranteed.
OUTPUT
Voltage Range:
Current Range:
Voltage Setting:
Current Setting:
Operating Mode:
Output Switch:
Output Terminals:
Output Protection:
Load Regulation:
Line Regulation:
Ripple & Noise
(20MHz bandwidth):
Transient Response:
Voltage Programming
speed:
Temperature Coefficient:
Status Indication:
METER SPECIFICATIONS
Meter Types:
Meter Resolutions:
Meter Accuracies:
MEMORY
The power supply saves the voltage, current and output-enable status at power down and
restores the settings at power up.
0V to 35V minimum
0.01A to 5A minimum
By coarse and fine rotary encoders or RS-232 interface.
Resolution 10mV. Setting accuracy 0.3% ± 20mV (for V >0.1V)
By single rotary encoder or RS-232 interface.
Resolution 10mA. Setting accuracy 0.6% ± 20mA (for I >0.1A)
Constant voltage (CV) or constant current (CC) with automatic cross-
over.
Electronic. Preset voltage and current displayed when off.
Universal 4mm safety binding posts on 19mm (0.75") pitch.
Output will withstand up to 40V forward voltage.
Reverse protection by diode clamp for reverse currents up to 3A.
<0.01% of maximum output for a 90% load change.
<0.01% of maximum output for a 10% line change
Typically <2mVrms, <10mV pk-pk, constant voltage mode.
<200µs to within 50mV of set level for a 5% to 95% load change.
Maximum time required for the output to settle within 1% of its total
excursion (for a resistive load). Excludes command processing time.
90% load
Up
210ms
Typically <100ppm/°C
Output on lamp. CC mode lamp.
Separate 4 digit meter for voltage and 3 digit meter for current with
14mm (0.56") LED displays. Reading rate 4/sec.
100mV, 10mA
Note that in constant voltage mode the meter will show the set
voltage to resolution of 10mV. However, the metering resolution is
limited to 100mV and when in CC mode the last digit will be set to
zero.
Voltage (CC mode) 0.3% ± 100mV.
Current (CV mode) 0.6% ± 20mA

Specification

No load
110ms
Down
90% load
No load
15ms
1.8s