the cable, generously spray disinfectant solution on a clean
towel and wipe the cable and connector. Discard used towel. To
disinfect the system, generously spray disinfectant on a clean
towel and wipe the cable and connector. Allow the disinfectant
solution to air dry.
3.
Clean and disinfect all surfaces of the dental treatment control
head according to the manufacturer's instructions.
4.
Connect a sterilized Steri-Mate
cable connector.
5.
Hold the handpiece (without an insert installed) over a sink
or drain. Activate the foot control and flush water through
handpiece for at least thirty seconds.
6.
Place a sterilized insert into the handpiece.
Shut down at the end of the day
®
1.
Flush the Cavitron
ultrasonic system according to the dental
treatment unit manufacturer's instructions.
2.
Remove any ultrasonic inserts and the Steri-Mate
Clean and sterilize the handpiece and all inserts used during the
procedure. (Refer to the Infection Control Information booklet and
the Cavitron Ultrasonic Inserts Directions for Use for cleaning and
sterilizing instructions.)
3.
Clean and disinfect the handpiece cable assembly by applying
a water-based disinfectant solution*, carefully following the
instructions provided by the solution manufacturer. To clean the
cable, generously spray disinfectant solution on a clean towel and
wipe the cable and connector. Discard used towel. To disinfect the
system, generously spray disinfectant on a clean towel and wipe
the cable and connector. Allow the disinfectant solution to air dry.
ENGLISH • 14
®
handpiece onto its mating
®
handpiece.
4.
Clean and disinfect all surfaces of the dental treatment
control head according to the manufacturer's instructions.
5.
Place the cable connector into the handpiece holder for
storage.
6.
Turn OFF the power and water to the dental treatment unit.
*NOTE: Water-based disinfection solutions are preferred. Some
alcohol-based disinfectant solutions may discolor
the handpiece cable and connector.