Errors and remedying them
10. Errors and remedying them
Error messages appear in the "Time display" field and are accompanied by an
audible signal. A distinction is made between operating errors and appliance
faults.
10.1
Operating errors
Incorrect head position
•
If the client is not sitting in the correct position under the Infrared heater,
model S32 or S33, the "CP" (Check Position) error message will appear in
the "Time display" field.
•
Check and correct the appliance position (middle of client's ear = middle of
side heater) and confirm using the corresponding illuminated Start button
(see "Temperature sensor/Client position" on page 33).
10.2
Appliance faults
Defective heating
•
The heating does not work properly.
There is a danger of overheating.
-
The message "E0" appears in field.
•
Switch off the appliance immediately, disconnect the mains plug, and noti-
fy the technical Customer Service Centre.
Defective temperature measurement device
•
Defective sensor or sensor line.
•
One or more essential functions of the appliance are defective.
-
The message "E1" appears in field
•
Switch off the appliance immediately, disconnect the mains plug, and noti-
fy the technical Customer Service Centre.
GB