Query The Operating Hours, Centrifugation Runs And Cycle Counter; Query The System Information; Immediate Display Of Centrifugation Data After Switch-On - Hettich ROTANTA 460 Robotic Mode D'emploi

Masquer les pouces Voir aussi pour ROTANTA 460 Robotic:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN
32

Query the operating hours, centrifugation runs and cycle counter

The operating hours are divided up into internal and external operating hours.
Internal operating hours: Total time the device was switched on.
External operating hours: Total time of previous centrifugation runs.
The query is performed as follows when the rotor is at a standstill:
The query can be made via the operating unit (key position "0" or "LOCK 2") or via the interface (key position
"LOCK 2"). The functions of the various key positions are described in the "Key-operated switch" chapter.
You can scroll backwards through the menu by pressing the
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.
Press the
key as often as necessary until the -> Operating Time is displayed.
PROG
Press the
key. The external operating hours are displayed, e.g. OP Time ext = 0h25m .
START
Press the
key. The internal operating hours are displayed, e.g. OP Time int = 1h36m .
PROG
Press the
key. The number of all centrifugation runs is displayed, e.g. Number of Starts = 10 .
PROG
Press the
key. The number of rotor cycles (centrifugation runs) of the current rotor code since the last time
PROG
the cycle counter was reset to "0" and the permitted number of rotor cycles are displayed, e.g. CYCLES = 5120
of 30000.
Press the
key. The number of all rotor cycles (centrifugation runs) of the current rotor code is displayed,
PROG
e.g. Rotor cycles total = 37490 . This value cannot be set.
Press the
key twice to exit the "Operating Time" menu or press the
OPEN / STOP
exit the "Machine Menu".
33

Query the system information

The following system information can be queried:
Centrifuge model
Mains voltage
Rotor information
Program version of the centrifuge
Program version of the frequency converter
The system information can be queried as follows when the rotor is at a standstill:
The query can be made via the operating unit (key position "0" or "LOCK 2") or via the interface (key position
"LOCK 2"). The functions of the various key positions are described in the "Key-operated switch" chapter.
You can scroll backwards through the menu by pressing the
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.
Press the
key as often as necessary until the ->Info is displayed.
PROG
Press the
key. The centrifuge model is displayed.
START
Press the
key. The mains voltage is displayed, e.g. Mains Voltage : 230 V .
PROG
Press the
key. The rotor code (Rotor), the maximum speed of the rotor (Nmax) and a centrifuging radius
PROG
(R) of the last rotor detected by the rotor detector are displayed. e.g.
Rotor 4 ∗ : Nmax = 4500 R=184 .
The last recognized rotor is marked with a star (∗). Use the adjusting knob
permitted in the centrifuge displayed.
The required centrifuging radius must be set according to the used accessories; see chapter "Entering
centrifugation parameters".
Press the
key. The program version of the centrifuge is displayed, e.g. SW version = V01.00.
PROG
Press the
key. The program version of the frequency converter is displayed e.g. FC-SW-Version = 4 .
PROG
Press the
key twice to exit the "Info" menu or press the
OPEN / STOP
"Machine Menu".
34

Immediate display of centrifugation data after switch-on.

Switch on the power supply. (Switch setting Ι).
The first time the display changes optically, (inverse display) press any key and keep it held down. The
centrifugation data is displayed immediately.
84/179
key.
T/°C
key.
T/°C
OPEN / STOP
Rev. 12 / 04.2020
key three times to
OPEN / STOP
to have the information of the rotors
key three times to exit the
AB5680DEENFRIT