Transport and storage

7.1

Transport

Before transporting the device, the transport securing device must be installed.
When the device and accessories are transported, the following ambient conditions must be complied with:
Ambient temperature: -20°C to +60°C
Relative humidity: 10% to 80%, non-condensing
7.2

Storage

The device and the accessories may only be stored in closed and dry rooms.
When the device and accessories are stored, the following ambient conditions must be complied with:
Ambient temperature: -20°C to +60°C
Relative humidity: 10% to 80%, non-condensing
8

Unpacking the centrifuge

Lift the carton upward and remove the padding.
Do not lift by the front panel.
Observe the weight of the centrifuge, refer to chapter "Technical specifications".
With a suitable number of helpers, hold the centrifuge on both sides and lift down from the pallet.
9

Initial operation

Remove the transport safety device (see instruction sheet on "Moving the equipment safely").
Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
If the centrifuge will be built into another machine or will be integrated in a system the
manufacturer of the complete system is responsible for its security.
Ventilation openings may not be blocked.
A distance of 300 mm must be maintained from the ventilation slots and openings of the centrifuge.
Centrifuge with RS232 interface:
Connect the RS232 centrifuge interface with an RS232 connecting cable (included in the scope of delivery for
certain models) to the PC.
Centrifuge with optical interface:
Connect the optical centrifuge interface with a light-conducting cable (included in the scope of delivery for certain
models) to the PC.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the power cord to a standard mains socket. For connection ratings refer to Chapter
"Technical specifications".
Switch on the power switch (switch position "Ι"). The LED's in the keys will flash.
The following displays appear one after the other:
1. The centrifuge model
2. The program version
3. The rotor code (Rotor), the maximum speed of the rotor (Nmax) and a centrifuging radius (R) of the last rotor
detected by the rotor detector.
4. The centrifugation data of the last used program or program 1.
62/179
Rev. 12 / 04.2020
AB5680DEENFRIT