Relative centrifugal force (RCF) and centrifuging radius (RAD)

Relative centrifugal force (RCF) depends on the centrifugal radius (RAD). Before RCF is set, the centrifugal
radius has to be set.
Press the
key as often as is required until the parameters RAD and RCF are displayed and the value of
RCF
the parameter, RAD is displayed in parentheses, 〈〉 e.g. RAD = 〈 146 〉
Use the adjusting knob
By changing the centrifuging radius, the value adjusts automatically to the RCF.
Press the
key again. The value of the parameter, RCF is displayed in 〈〉 parentheses, e.g. RAD = 146
RCF
RCF = 〈 3695 〉.
Press the
key to save the set RCF value. See the "Inputting or changing programs" chapter.
PROG
Press the
or
RCF
START
Only by storing (STO) the set RCF value will the RPM value resulting from that be accepted.
18.5

Start-up and run-down parameters

The set start-up and run-down parameters are displayed.
x: 1-9 = start-up stage, t = start-up time
y: 1-9 = braking stage, 0 = unbraked run-down, t = run-down time
y
x
18.5.1

Start-up stage and start-up time

It is only possible to set start-up times if these have been activated. See the chapter "Activating or deactivating
start-up and run-down times".
Press the
key. The parameter
= start-up stage,
t
= start-up time
Press the
key to switch between the start-up stage and start-up time.
TIME
Set the desired stage or time with the rotary knob .
If necessary, press the
To apply the setting to the display, either press the
the centrifugation data are displayed.
18.5.2

Braking stage and run-down time

No B-brake stages can be set for this device. The activation of the B-brake stages in the "Settings" menu is
not possible (parameter B-Ramp = off ). B-brake stages are similar to an exponential braking curve.
It is only possible to set run-down times if these have been activated. See the chapter "Activating or
deactivating start-up and run-down times".
Press the
key until the parameter
= braking stage,
t
= run-down time
Press the
key to switch between the braking stage and run-down time.
TIME
Set the desired stage or time with the rotary knob .
If necessary, press the
To apply the setting to the display, either press the
the centrifugation data are displayed.
18.5.3

Brake switch-off speed

Press the
key as often as necessary until the parameter N Brake is shown.
Use the adjusting knob
Press the
or
START
72/179
to set the centrifuging radius you want.
keys to apply the setting to the display.
or
t
is displayed.
key to set the next parameter.
START
or
t
is displayed.
key to set the next parameter.
START
to set the value you want.
keys to apply the setting to the display.
Rev. 12 / 04.2020
RCF = 3695 . The LED is lit in the key.
key or press the
key as often as is required until
key or press the
key as often as is required until