EN
15.1
Running time
In order to set continual running, the minutes, seconds and hours have to be set to zero.
Continual running is represented in the display by the following symbol, "".
key. The parameters t/hms is displayed. The minutes (m) are shown in parentheses , and can
Press the
TIME
be changed.
Use the adjusting knob
key. The seconds (s) are shown in parentheses and can be changed.
Press the
TIME
Use the adjusting knob
key. The hours (h) are shown in parentheses and can be changed.
Press the
TIME
Use the adjusting knob
To apply the setting to the display, either press the
the centrifugation data are displayed.
15.2
Starting runtime timing
The start to runtime timing can only be set if the function, "Timing begins at speed" has been activated. See
chapter "Activate or deactivate function "Timing begins at speed"". Ex works, the function is deactivated.
Press the
key as often as is required until Timing begins at start or Timing begins at speed is displayed.
TIME
Use the knob
to set Timing begins at start or Timing begins at speed.
Timing begins at Start = runtime timing begins immediately after the centrifugation run has been started.
Timing begins at Speed = runtime timing begins once the set speed has been reached.
If Timing begins at speed is set, this is shown in the display by the
Press the
or
TIME
START
15.3
Speed (RPM)
Press the
key. The parameters RPM is displayed.
RPM
Use the adjusting knob
Press the
or
RPM
START
15.4
Relative centrifugal force (RCF) and centrifuging radius (RAD)
Relative centrifugal force (RCF) depends on the centrifugal radius (RAD). Before RCF is set, the centrifugal
radius has to be set.
Press the
key as often as is required until the parameters RAD and RCF are displayed and the value of
RCF
the parameter, RAD is displayed in parentheses, e.g. RAD = 146
Use the adjusting knob
By changing the centrifuging radius, the value adjusts automatically to the RCF.
Press the key again. The value of the parameter, RCF is displayed in parentheses, e.g. RAD = 146
3695.
Use the adjusting knob
Press the
or
RCF
START
52/165
to set the value you want.
to set the value you want.
to set the value you want.
START
keys to apply the setting to the display.
to set the value you want.
keys to apply the setting to the display
to set the centrifuging radius you want.
to set the RCF you want.
keys to apply the setting to the display.
key or press the
key as often as is required until
TIME
icon, on the left, next to the time.
RCF = 3695. The LED is lit in the key.
RCF =