Reset Cycle Counter To "0" And Enter The Maximum Permitted Number Of Cycles - Hettich ROTANTA 460 Robotic Mode D'emploi

Masquer les pouces Voir aussi pour ROTANTA 460 Robotic:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

24.2

Reset cycle counter to "0" and enter the maximum permitted number of cycles

If the rotor is at standstill, this can be set as follows:
You can scroll backwards through the menu by pressing the
The operation can be cancelled at any time by pressing the
stored.
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.
Press the
key as often as necessary until the ->Operating Time is displayed.
PROG
Press the
key. The external operating hours are displayed, e.g. OP Time ext = 0h25m .
START
Press the
key as often as is required until the rotor cycles are displayed, e.g. Cycles = 30001 of 30000 .
PROG
key. The number of rotor cycles is displayed in parentheses áñ , e.g. Cycles = 〈 30001 〉 of
Press the
RCF
30000 .
Turn the knob
to the left to reset the number of rotor cycles to "0".
If the rotor cycles are not reset to "0", press the
to be displayed and the setting is not saved.
Press the
key. The maximum permitted number of rotor cycles is displayed in parentheses 〈〉, e.g. Cycles
RCF
= 0 of 〈 30000 〉.
Use the knob
to set the maximum permitted number of rotor cycles specified on the hanger.
Press the
key to save the settings.
START
In confirmation, Store cycles ... is displayed briefly. Next the rotor cycles are displayed, e.g. Cycles = 0 of
30000 .
Press the
key twice to exit the "Operating Time" menu or press the
OPEN / STOP
exit the "Machine Menu".
AB5680DEENFRIT
key.
T/°C
OPEN / STOP
key, which causes Max cycles 〈 = actual cycles
START
Rev. 12 / 04.2020
key. In this case, the settings are not
key three times to
OPEN / STOP
EN