42.2.4
Trunnions
With swing-out rotors the trunnions must be regularly lubricated (Hettich Lubricating Grease No. 4051) in order to
ensure consistent swinging out of the hangers.
42.2.5
Rotors and accessories with limited service lives
The use of certain rotors, hangers and accessory parts is limited by time.
These are marked with the maximum permitted number of operating cycles or with an expiration date and the
maximum permitted number of operating cycles or just with the expiration date; e.g.:
- "einsetzbar bis Ende: Ι V. Quartal 2011 / usable until end of: Ι V. Quarter 2011" or
"einsetzbar bis Ende Monat/Jahr: 10/2011 / usable until end of month/year: 10/2011"
- "Max. Lauf Zyklen / max. cycles: 40000".
For safety reasons, rotors, hangers and accessory parts may no longer be used if either the indicated
maximum number of operating cycles or the indicated expiration date has been reached.
The number of centrifuge operations can be queried; see the chapter on "Querying the operating hours and the
number of centrifuge operations".
42.3
Autoclaving
The following accessory can be autoclaved at 121°C / 250°F (20 min):
•
Swing-out rotors
•
Hanger made of metal
•
Stands
•
Reductions
Otherwise you must ask the manufacturer.
No statement can be made about the degree of sterility.
Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics.
42.4
Centrifuge containers
•
With leakiness or after the breakage of centrifuging containers broken container parts, glass splinters and leaked
centrifugation material are to be completely removed.
•
The rubber inserts as well as the plastic sleeves of the rotors are to be replaced after a glass breakage.
Remaining glass splinters cause further glass breakage!
•
If this concerns infectious material, a disinfection process is to be executed immediately.