Figure 4 Contents of the application box
1 KF standard beaker
®
2 KF white Teflon
conical adapter
3 Magnetic stir bars (10x)
4 USB applications key
5 Sensor
1
2 filled with a molecular sieve, and 1 empty

Installation

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Installation guidelines
• This instrument is for indoor use only.
• The power supply connector on the rear panel must be easily accessible so the power can be
disconnected quickly in case of emergency.
• Keep the instrument away from temperature extremes, including heaters, direct sunlight and other
heat sources.
• Put the instrument on a stable and level surface in a well ventilated place.
• Make sure that there is at least 15 cm (6 in.) of space on all sides of the instrument to prevent
electrical parts from overheating.
• Do not operate or keep the instrument in dusty, damp or wet locations.
• Always keep the surface of the instrument and all accessories dry and clean.
6 Syringe holding ring
7 Syringe
8 Glass bottle
9 Bottle caps (3 x GL45)
10 Desiccant cartridges (3x)
C A U T I O N
1
English 7