Hach TitraLab KF1000 Série Manuel D'utilisation De Base page 10

Masquer les pouces Voir aussi pour TitraLab KF1000 Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 23
Install the sensor holder accessories
Add the stir bar to the beaker, and then attach the beaker to the sensor holder. Install the desiccant
cartridge and the sample stopper. Refer to the illustrated steps that follow.
Prepare the tubes
Remove any bends in the end of the tubes. Refer to the illustrated steps that follow.
Connect the tubes
Arrow symbols identify the inlet and outlet ports for the syringe and the pump connections. The "up"
arrow is the outlet port. The "down" arrow is the inlet port. Turn the tube connectors on the inlet and
outlet ports of the syringe and pump until they click.
The suction tube that drains into the waste bottle must be installed on the bottom of the beaker for
good drainage.
The diffusion tip and tube holder on the outlet tube, are pre-installed at the optimal positions. Do not change the
position of the diffusion tip or the tube holder.
Refer to the illustrated steps that follow.
10 English
N O T I C E

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Titralab kf1121

Table des Matières