Hach CL17sc Manuel D'utilisation page 3

Masquer les pouces Voir aussi pour CL17sc:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
Table of Contents
1 Specifications
on page 3
2 General information
on page 4
3 Installation
on page 8
4 Configuration
on page 17
5 Operation
on page 18
Section 1 Specifications
Specifications are subject to change without notice.
Specification
Dimensions (W x H x D)
Enclosure
Shipping weight
Mounting
Protection class
Pollution degree
Electrical installation category
Power requirements
Operating temperature
Operating humidity
Storage temperature
Fittings
Indicator lights
Certifications
Warranty
Specification Details
Pressure
0.31 to 5.17 bar (4.5 to 75 psig) supplied to Y-strainer; 0.1 to 0.34 bar (1.5 to 5 psig) supplied to
analyzer
Flow rate
60 to 200 mL/min through the instrument (measured at analyzer drain)
Temperature
5 to 40 °C (41 to 104 °F)
Filtration
Y-strainer with 40 mesh screen or higher
1
Drain holes closed with plugs for testing requirements.
6 Calibration and adjustment
7 Maintenance
8 Troubleshooting
9 Replacement parts and accessories
Table 1 General specifications
Details
32.9 x 34.2 x 17.7 cm (12.9 x 13.5 x 7.0 in.)
IP66 per IEC 60529 with the door closed and latched
4.1 kg (9 lb) without bottles; 5.1 kg (11.2 lb) with full bottles
Wall mount
III
3
I (indoor installation)
12 VDC, 400 mA maximum (supplied by the controller)
5 to 40 °C (41 to 104 °F)
0 to 90% non-condensing relative humidity
-40 to 60 °C (-40 to 140 °F)
Sample line: ¼-in. OD quick-connect fitting for plastic tubing
Drain lines: slip-on fitting for ½-in. ID soft plastic tubing
Analyzer status and measurement cycle
CE DoC, FCC SDoC, IC, RCM, KC
1 year (EU: 2 years)
Table 2 Sample requirements
on page 18
on page 18
on page 24
on page 25
1