Cleaning Of Accessories; Cleaning The Main Unit - Gima TOBI Mode D'emploi

Masquer les pouces Voir aussi pour TOBI:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

To clean the plastic housing of the device wear disposable latex gloves
Washing and / or cleaning the autoclavable jar as to be carried out as follows:
Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances;
Disconnect the jar from the device
Disconnect all tubes from the jar and the protection filter
Empty and dispose of the content and of the suction catheter according to the laws in force in your country;
Separate all parts of the cover (overflow valve, o-ring);
After disposing of disposable parts and disassembling the jar wash in running cold water and rinse thouroughly. Then soak in warm water
(temperature shall not exceed 60°C). Wash thouroughly and if necessary use a non-abrasive brush to remove incrustations.
Rinse in running warm water and dry all parts with a soft cloth (non-abrasive). The jar and the cover can be autoclaved by placing the parts into the
autoclave and running one sterilization stem cycle at 121°C (1 bar relative pressure – 15min) making sure that the jar is positioned upsidedown.
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1).
Beyond this limit the physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended.
After sterilization and cooling at environment temperature of the parts make sure that these are not damaged. Assemble the jar as follows:
Place the overflow valve into its seat in the cover (under VACUUM connector)
Insert floating valve keeping the o-ring towards the opening of the cage
Place the o-ring into its seat around the cover
After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leackages or liquid exit
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 120°C (1 bar relative pressure – 15min). The conical connector can
be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15min). The device is ready for a new employment now.
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
Available under request with different versions with complete jar 2000cc.
Antibacterial Filter: The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the pneumatic circuit.
The filter is for a single patient use which will protect patients and machines from cross contamination.
When the filter is wet, it's not possible to use the unit therefore the filter should be changed immediately.
In case of possible contamination or discolouration, change the filter immediately. Don't use the suction unit without the protection filter fitted.
Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use. Reuse may cause cross-infections.
Don't use after lapse of the sell-by date
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standards
on material biocompatibility.
To clean the plastic housing of the device wear disposable latex gloves. To clean the device external parts always use a cotton cloth dampened
with detergent. Don't use abrasive or solvent detergents.
PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE EQUIPMENT DO NOT GET IN TOUCH
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glasses) to avoid getting in contact with
contaminating substances (after each utilization cycle of the machine).

CLEANING OF ACCESSORIES

ACCESSORIES SUPPLIES
DESCRIPTION
COMPLETE ASPIRATION JAR 1000cc
CONICAL FITTING
TUBES SET 6mm x 10mm (TRASPARENT SILICON)
ASPIRATION PROBE CH 20
ANTIBACTERIAL FILTER

CLEANING THE MAIN UNIT

WITH LIQUIDS. NEVER CLEAN THE EQUIPMENT WITH WATER.