5 SCPI Programming Reference
LIST causes the output to follow the list values when a trigger occurs.
ARB causes the output to follow the arbitrary waveform values when a trigger occurs.
Parameter
FIXed|STEP|LIST|ARB, *RST FIXed
Sets the power mode to Step: POW:MODE STEP
[SOURce:]POWer:SLEW[:IMMediate] <value>|MIN|MAX|INFinity
[SOURce:]POWer:SLEW[:IMMediate]? [MIN|MAX]
Sets the power slew rate. The slew rate is set in watts per second and affects all programmed power
changes, including those due to the output state turning on or off. The slew rate can be set to any
value between 0 and 9.9E+37. For very large values, the slew rate will be limited by the unit's listed
programming speed and bandwidth. Keywords MAX or INFinity set the slew rate to maximum.
Parameter
0 – 9.9E+37, *RST MAX
Sets the output slew rate to 10 W per second: POW:SLEW 10
The query returns the value that was sent. If the value is less than the minimum slew rate, the min-
l
imum value is returned. The resolution of the slew setting is the same as the minimum value. This
can be queried using POWer:SLEW? MIN.
[SOURce:]POWer:SLEW:MAXimum 0|OFF|1|ON
[SOURce:]POWer:SLEW:MAXimum?
Enables or disables the maximum slew rate override. When enabled, the slew rate is set to its
maximum value. When disabled, the slew rate is set to the immediate value set by the POWer:SLEW
command. Use POWer:SLEW? MAX to query the maximum slew rate that was set.
Parameter
0|OFF|1|ON, *RST ON
Enable the maximum slew rate: POW:SLEW:MAX ON
POWer:SLEW:MAX is coupled to POWer:SLEW. If POWer:SLEW sets the rate to MAX or INFinity,
l
POWer:SLEW:MAX is enabled. For any other slew rate value, POWer:SLEW:MAX is disabled.
228
Typical Return
FIX, STEP, LIST, or ARB
Typical Return
<max value>
Typical Return
0 or 1
Keysight RP7970 & RP7980 Series Operating and Service Guide