Error which has
occurred
Possible causes
The measuring signals
could not, or not correctly,
be read. This can be caused
by incorrect application of
the cuff, moving, talking or
by a very weak pulse.
Cuff does not infl ate or
does not infl ate fast
enough. This can, among
other things, be due to a
too loosely applied cuff
or to movement.
Air release during the
measuring procedure is
too fast or too slow. The
cuff could have become
undone or loo sened itself.
A movement during the
measu ring procedure is
also a possibility.
The pressure in cuff
exceeds 300 mmHg.
Therefore, an automatic
fall in pressure occurs.
Remedy
Check the correct position
of the cuff. Do not talk or
move during the measur-
ing procedure. Also observe
the 10 golden rules as well
as the notes in the box in
chapter 3.3.
Apply the cuff so that you
are able to push about
two fi ngers between the
cuff and the upper arm.
Air tube is not correctly
inserted into the device.
Check the correct position
of the connector. If this
error occurs often, use a
new cuff.
Check the correct position
of the cuff. Do not move
during the measuring
procedure.
Please rest for at least a
minute and take the
measurement again.