Area 3:
If applicable, data specified by the manufacturer are contained in this area:
These specify
- the transducer type,
- the measured quantity,
- the electrical output signal,
- the required excitation.
- the measured sensitivity
Area 4:
The actual user can modify the last of these areas with, for instance:
- a short comment in text form,
- filter settings,
- zero value
Example:
The TEDS content of sensor P3 Top Class BlueLine/5000 bar with ident. no.
111410256, manufactured in March 2007
TEDS
Manufacturer
Model
Version letter
Version number
Serial number
Template: Bridge Sensor, Full precision
Transducer Electrical Signal Type
Minimum Pressure
Maximum Pressure
Minimum Electrical Value
Maximum Electrical Value
Mapping Method
Bridge type
Impedance of each bridge element
Response Time
Excitation Level (Nominal)
Excitation Level (Minimum)
Excitation Level (Maximum)
Calibration Date
Calibration Initials
Calibration Period (Days)
Measurement location ID
HBM
HBM
P3
11
1410256
Bridge Sensor
0.000
500M
0.000
0.999m
Linear
Full
350
0
5.0
0.5
7.5
4-Mrz-2007
HBM
730
0
BlueLine
Pa
Pa
V/V
V/V
Ohm
sec
V
V
V
days
A2222-4.0 en/de/fr