Chapter 5
TROUBLESHOOTING
Trouble
Can not turn
on the device
No blood pressure
result
No SpO
result
2
No TEMP result
Possible reason
The built-in battery is drained
Battery is not installed
Some parts provided by others are inserted to
the connector
The cuff is wrapped around the arm incorrectly
The windpipe is not well inserted to NIBP jack
SpO
probe is not plugged
2
to Port1 or Port2
Temperature probe is not well plugged
to Port1 or Port2
Make measurement before "READY" appears
on the temperature probe screen
ENGLISH
Solution
Recharge by connecting the power
supply adapter
Install the Lithium battery
Remove the related parts and try again.
Wrap the cuff around the arm correctly
Insert the windpipe to the NIBP jack
Plug SpO
probe to Port1 or Port2
2
Plug temperatur2 probe to Port1 or Port2
Do not make measurement until
"READY" appears on the temperature
probe screen