with mark of "PORT1" or "PORT2". When LCD screen displays
the probe is connected successfully.
2. When the screen shows as figure 2.7B and the temperature unit °C is blinking, then
the user can start to take measurement.
3. Insert the measuring tip of temperature probe into the earhole and press the measuring key
to start the measurement. A short beep means the measurement is finished and the result
will be displayed on the screen.
Note:
� If the temperature probe detects a hardware failure, the display screen on the infrared
temperature probe will show "Err" and will not enter into measuring mode.
� The infrared temperature probe will stand by automatically if no operation for 1 min. If you
need to make a measurement once more, please press the measuring key and repeat step
2 and step 3.
Normal body temperature varies in a range. The following table shows the temperature varying
range at different body position, so it is meaningless to simply compare the temperature rea-
dings from different position.
Temperature varying range at different body positions
Arm
Oral
Rectal
Ear
ENGLISH
34.7 ~ 37.3°C
35.5 ~ 37.5°C
36.6 ~ 38.0°C
35.8 ~ 38.0°C