Gima MONITEUR SPOT-CHECK PC-300 Manuel De L'utilisateur page 57

Masquer les pouces Voir aussi pour MONITEUR SPOT-CHECK PC-300:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

57
 When the user is connected with other devices, the total leakage current may exceed
the limitation and cause potential danger to the user as a result.
 All combinations of equipment must be in compliance with standard of IEC 60601-1-1
medical and electric system requirements.
 Although biocompatibility tests have been performed on all the applied parts, some
exceptional allergic patients may still have anaphylaxis. Do NOT apply to those who have
anaphylaxis.
 All the connecting cables and rubber tubes of the applied parts should be kept away
from the patient' neck to prevent any possible suffocation of the patient.
 All the parts of the monitor should NOT be replaced at will. If necessary, please use
the components provided by the manufacturer or those that are of the same model and
specifications as the accessories along with the monitor which are provided by the same
factory, otherwise, negative effects concerning safety and biocompatibility etc. may be
caused.
 If the monitor falls off accidentally, please do NOT operate it before its safety and technical
performance have been tested minutely and positive testing results obtained.
 Do NOT open the device cover without authorization. The cover should only be opened
by qualified service personnel.
 When disposing of the monitor and its accessories, the local law should be followed.
ENGLISH

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Moniteur spot-check pc-301Moniteur spot-check pc-30235162

Table des Matières