Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied.
Note 2: These guide lines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a) Field intensity from fixed transmitters such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
ENGLISH
Where P is the maximum nominal output
voltage of the transmitter in Watt (W) depending
on the manufacturer of the transmitter and d
is the recommended separation distance
in meters (m).
The intensity of the field from the fixed RF
transmitters, as determined by an
electro-magnetic study of the site (a) could
be lower than the level of conformity of each
frequency interval (b). It is possible to check
for interference in proximity to devices
identified by the following symbol: