Guidance and manufacturer's declaration – electromagnetic emissions
The ME equipment is intended for use in the electromagnetic environment specifi ed
below. The customer or the user of the ME equipment should assure that
it is used in such an environment.
Emissions Test
Compliance
The ME equipment uses RF energy only for its internal function.
RF Emissions
Group 1
CISPR 11
RF Emissions
Class B
CISPR 11
Harmonic emissions
Not Applicable
IEC 61000-3-2
Voltage fl uctuations/
Not Applicable
fl icker emissions
Non-Life Support Equipment Separation Distance Calculation
(3Vrms / 3V/m compliance)
150 kHz to 80 MHz
in ISM bands
3,5
Rated maximum output
d = [
] P
V
power of transmitter (W)
1
0.01
0.12
0.1
0.37
1
1.17
10
3.69
100
11.67
14
All manuals and user guides at all-guides.com
Electromagnetic environment – guidance
Therefore, its RF emissions are very low and are not likely to
cause any
interference in nearby electronic equipment.
Complies
The ME equipment is
solely battery powered.
Separation distance according to
frequency of transmitter (m)
80 MHz to
800 MHz to
800 MHz
2.5 GHz
3,5
7
d = [
] P
d = [
] P
E
E
1
1
0.12
0.23
0.37
0.74
1.17
2.33
3.69
7.38
11.67
23.33
Guidance and manufacturer's declaration – electromagnetic immunity
The ME is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the ME should assure that it is used in such an environment.
IEC60601 test
Compliance
Immunity test
level
Electrostatic
±6kV Contact
discharge (ESD)
Complies
±8kV Air
IEC 61000-4-2
Radiated RF
3 V/m 80MHz to
Complies
2.5GHz
IEC 61000-4-3
Conducted RF
3Vrms 150kHz to
Applicable
80MHz
(no electrical
IEC 61000-4-6
cabling)
Electrical fast
±2kV
transient
power line
Applicable
IEC 61000-4-4
±1kV I/O lines
±1kV
Surge
diff erential
Applicable
IEC 61000-4-5
±2kV common
Power frequency
magnetic fi eld
3 A/m
Complies
IEC 61000-4-8
Voltage dips, short
>95% dip 0.5 cycle
interruptions and
voltage variations
60% dip 5 cycles
on power supply
Applicable
70% dip 25 cycles
input lines
95% dip 5 sec.
IEC 61000-4-11
Electromagnetic environment– guidance
level
Floors should be wood, concrete, or ceramic
tile. If fl oors are covered with synthetic mate-
rial, the relative
humidity should be at least 30%
Field strengths outside the shielded location
from fi xed RF transmitters, as determined by
an electromagnetic site survey, should be less
than 3 V/m.
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
Not
Separation distance
calculation provided above. If a known trans-
mitter is present the specifi c distance can be
calculated using the equations.
Not
The ME equipment
is solely battery powered.
Not
Power frequency
magnetic fi elds should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
The ME equipment
Not
is solely battery
powered.
15