Resetting the Patient
Run Hours

Settings Adjustment

1. When the pump is in Standby, press and hold
the Right Control button for 2 seconds. The
Patient Run Hours value will flash for 2 minutes.
Press and hold the Middle Control button for 3
seconds to reset the Patient Run Hours to zero.
This is the pump run time since the patient run
hours was last reset.
2. Press the Right Control button again to save the
new setting. If there is no action within the 2
minutes of flashing, the selected setting will be
saved.
The pump is configured to give the recommended
therapy for each garment type and does not require
any direct setting by the clinician or nurse.
If the physician requires different therapy settings for
uniform calf and/or calf & thigh garments, then limited
changes can be made to the pump pressure setting.
These changes, and returning the pump to the
default settings, can only be made either:
• By contacting service personnel through your
local Arjo sales office, or
• By trained authorised technical personnel at the
facility.
The pump pressures for foot garments and
sequential calf and calf & thigh garments are
fixed and cannot be changed.
The pressure range and factory default pressure
for the uniform calf and calf & thigh garments are
detailed in the Pressure Range on page 29.