Warnings/Alarms On detection of a fault condition, the pump provides
To Change the Audible
Alarm Volume Setting
a visual-only warning followed by an audible and
visual alarm if the fault is not cleared.
1. The visual-only warning is a fault icon on the
LCD screen. The LED indicators remain
unchanged and there is no audible alarm.
2. If the fault is not cleared then the warning is
replaced by an audible and visual alarm, which
consists of:
• A fault icon on the LCD screen.
• The LED indicators on the front and underside
of the pump change to yellow.
• An audible alarm will sound.
The warning and alarm can be cleared by either:
• Rectifying the fault on the system, or
• Pressing the Middle Control button to put the
pump into Standby.
Refer to Section 7, Page 22 "Troubleshooting" for
the alarms, their possible causes and their remedies.
1. When the pump is in Standby, press the Left
Control button for 2 seconds to enter Audible
Alarm Volume Setting mode.
2. The Audible Alarm Volume and Audible Alarm
Status icons will be displayed in the bottom left
of the screen.
Audible Alarm
Volume
3. Each time the Middle Control button is pressed
the volume setting increases; if the volume
setting is at maximum, pressing the button again
will cycle back to the minimum setting.
4. The pump will save the selected volume setting
when the Left Control button is pressed.
If the Left Control button is not pressed for more
than 2 minutes, the selected setting will be
stored, the Volume and Audible icons will be
removed and the pump will exit the Audible Alarm
Volume Setting mode.
Minimum
Volume
18
Audible Alarm
Status
Medium
Volume
Maximum
Volume