time of 27 seconds. The foot garment icon is
black while it is inflated.
3. The foot garment is deflated to zero. Both
garment icons are outlines.
4. Foot garment 2 is then inflated to the target
pressure, with an inflate time of 3 seconds and a
deflate time of 27 seconds. The foot garment
icon is black while it is inflated.
5. The foot garment is deflated to zero. Both
garment icons are outlines.
6. This cycle of alternate garment inflations repeats
from step 2 (above) until the therapy is stopped.
If calf (or calf & thigh) garments are attached to the
pump, then each of these garments is inflated to the
target pressure, with an inflate time of 12 seconds
and a deflate time of 48 seconds.
If a foot garment and a calf (or calf & thigh) garment
are attached to the pump, then since the inflate and
hold times for a foot garment are shorter than for a
calf (or calf & thigh) garment, there is a modified
inflation sequence: the foot garment is always
15