Télécharger Imprimer la page

HemoCue Glucose 201 Manuel D'utilisation page 33

Publicité

Symptom
Explanation
E35
1.
E70–E71
Erroneous kinetics.
1.
2.
ECU
1.
2.
HHH
1.
The current measurement has been incorrectly stored and
will remain as error code E35 until the entire memory has
been deleted.
The cuvette is faulty.
The sample might be grossly lipemic.
The cuvette is empty or not filled with enough sample. Please
note, the ECU function captures primarily empty cuvettes,
not under-filled cuvettes.
No chemical reaction is identified in the blood filled cuvette.
Measured value exceeds 27.8 mmol/L (500 mg/dL) for a
whole blood system and 30.8 mmol/L (555 mg/dL) for a
plasma equivalent system.
Action
1a.
No action is normally required.
1b.
If storage problems continue, the analyzer needs service. Contact your local
distributor.
1–2. Check that the system is used according to the instru ction for use.
Fill a NEW cuvette and perform a measurement.
If the error code appears again, use a suitable laboratory reference method to
analyze the specimen.
1–2. Fill a NEW cuvette and perform a measurement.
If the ECU function appears again, contact your local distributor or HemoCue
AB. Use another HemoCue analyzer or a suitable laboratory reference method to
analyze the specimen.
1.
The result is above the measuring range.
33

Publicité

loading

Produits Connexes pour HemoCue Glucose 201