b            With  a  frequency  range  over  150  kHz  to  80  MHz,  the  field  strengths  should  be  less  than  3  V/m.  
 
Recommended  distances  between  portable  and  mobile  RF  communications  equipment  
The  e-­scope®  is  intended  for  use  in  an  electromagnetic  environment  in  which  RF  emissions  
are  controlled.  The  customer  or  user  of  the  e-­scope®  can  help  to  avoid  electromagnetic  
interference  by  observing  the  minimum  distance  between  portable  and  mobile  RF  
communications  equipment  (transmitters)  and  the  e-­scope®  in  accordance  with  the  maximum  
output  power  of  the  communication  equipment.  
Separation  distance  according  to  the  frequency  of  the  transmitter    
Rated  maximum  
output  power  of  the  
transmitter  (W)  
0.01  
0.1  
1  
10  
100  
For  transmitters  with  a  maximum  output  power  not  listed  above,  the  recommended  distance  d  
in  metres  (m)  can  be  estimated  using  the  equation  for  the  transmitter  frequency,  where  P  is  the  
maximum  output  power  of  the  transmitter  in  watts  (W).  according  to  the  transmitter  
manufacturer.  
NOTE  1  At  80  MHz  and  800  MHz,  the  separation  distance  applies  to  the  higher  frequency  
range.  
NOTE  2  These  guidelines  may  not  apply  in  all  situations.  The  electromagnetic  propagation  is  
affected  by  absorption  and  reflection  of  structures,  objects  and  people.  
 
 
20
and  the  e-­scope®  
(m)  
150  KHz  to  80  MHz  
0.12  
0.12  
0.38  
0.38  
1.2  
1.2  
3.8  
3.8  
12  
12  
80  MHz  to  800  MHz  
800  MHz  to  2.7  GHz  
0.23  
0,23  
0.73  
0,73  
2.3  
7.3  
23  
 
2,3  
7,3  
23  
13  