b With a frequency range over 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.
Recommended distances between portable and mobile RF communications equipment
The e-scope® is intended for use in an electromagnetic environment in which RF emissions
are controlled. The customer or user of the e-scope® can help to avoid electromagnetic
interference by observing the minimum distance between portable and mobile RF
communications equipment (transmitters) and the e-scope® in accordance with the maximum
output power of the communication equipment.
Separation distance according to the frequency of the transmitter
Rated maximum
output power of the
transmitter (W)
0.01
0.1
1
10
100
For transmitters with a maximum output power not listed above, the recommended distance d
in metres (m) can be estimated using the equation for the transmitter frequency, where P is the
maximum output power of the transmitter in watts (W). according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance applies to the higher frequency
range.
NOTE 2 These guidelines may not apply in all situations. The electromagnetic propagation is
affected by absorption and reflection of structures, objects and people.
20
and the e-scope®
(m)
150 KHz to 80 MHz
0.12
0.12
0.38
0.38
1.2
1.2
3.8
3.8
12
12
80 MHz to 800 MHz
800 MHz to 2.7 GHz
0.23
0,23
0.73
0,73
2.3
7.3
23
2,3
7,3
23
13