Directives  and  manufacturer's  declaration  -­  Electromagnetic  immunity  
The  e-­scope®  instruments  are  intended  for  use  in  the  electromagnetic  environment  specified  below.  The  customer  or  
user  of  the  e-­scope®  should  ensure  that  it  is  used  in  such  an  environment.  
Immunity  testing  
IEC  60601  test  level  
 
Guided  RF  
Diturbances  acc.  
3  Vrms  
to  
0,5  MHz  bis  80MHz  
IEC61000-­4-­6  
6  V  in  ISM  frequency  
bands  between  0.15  
 
MHz  and  80  MHz    
 
80%  AM  at  1  kHz  
 
 
 
 
 
 
 
 
 
 
Radiated  RF  
3  V/m  
IEC  61000-­4-­3  
80  MHz  to  2.7  GHz  
 
 
Proximity  fields  
380  -­  390  MHz  
from  RF  wireless  
27  V/m;;  PM  50%;;  18  Hz  
communications  
430  -­  470  MHz  
equipment  
28  V/m;;  (FM  ±5  kHz,  1  
kHz  sine)  
PM;;  18  Hz11  
704  -­  787  MHz  
9  V/m;;  PM  50%;;  217  Hz  
800  -­  960  MHz  
28  V/m;;  PM  50%;;  18  Hz  
1700  -­  1990  MHz  
28  V/m;;  PM  50%;;  217  
Hz  
2400  -­  2570  MHz  
28  V/m;;  PM  50%;;  217  
Hz  
5100  -­  5800  MHz  
9  V/m;;  PM  50%;;  217  Hz  
NOTE  1  At  80  MHz  and  800  MHz,  the  higher  frequency  range  applies.  
NOTE  2  These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is  affected  by  absorption  and                                                        
reflection  from  structures,  objects  and  people.  
a              field  strengths  of  fixed  transmitters,  such.  B.  Base  stations  for  radio  (cellular  /  cordless)  telephones  and  land  
mobile  radios,  amateur  radio,  AM  and  FM  broadcasting  and  television  broadcasting  can  not  be  predicted  
theoretically  accurately.  To  assess  the  electromagnetic  environment  due  to  fixed  RF  transmitters,  an  
electromagnetic  examination  should  be  considered.  If  the  measured  field  strength  at  the  location  where  the  e-­
scope®  is  used  exceeds  the  above  RF  compliance  level,  the  e-­scope®  should  be  observed  to  verify  normal  
operation.  If  abnormal  performance  is  observed,  additional  measures  may  be  required,  such  as  reorienting  or  
shifting  the  e-­scope®.  
b            With  a  frequency  range  over  150  kHz  to  80  MHz,  the  field  strengths  should  be  less  than  3  V/m.  
 
 
Recommended  distances  between  portable  and  mobile  RF  communications  equipment  
The  e-­scope®  is  intended  for  use  in  an  electromagnetic  environment  in  which  RF  emissions  
are  controlled.  The  customer  or  user  of  the  e-­scope®  can  help  to  avoid  electromagnetic  
interference  by  observing  the  minimum  distance  between  portable  and  mobile  RF  
communications  equipment  (transmitters)  and  the  e-­scope®  in  accordance  with  the  maximum  
output  power  of  the  communication  equipment.  
Separation  distance  according  to  the  frequency  of  the  transmitter    
Rated  maximum  
output  power  of  the  
transmitter  (W)  
0.01  
0.1  
1  
10  
100  
For  transmitters  with  a  maximum  output  power  not  listed  above,  the  recommended  distance  d  
in  metres  (m)  can  be  estimated  using  the  equation  for  the  transmitter  frequency,  where  P  is  the  
maximum  output  power  of  the  transmitter  in  watts  (W).  according  to  the  transmitter  
manufacturer.  
NOTE  1  At  80  MHz  and  800  MHz,  the  separation  distance  applies  to  the  higher  frequency  
range.  
NOTE  2  These  guidelines  may  not  apply  in  all  situations.  The  electromagnetic  propagation  is  
affected  by  absorption  and  reflection  of  structures,  objects  and  people.  
 
Compliance    
Portable  and  mobile  RF  communications  equipment  
 
should  not  be  used  closer  to  any  part  of  the  non-­contact  
e-­scope®,  including  the  cables,  than  the  recommended  
Not  applicable  
distance,  which  is  calculated  using  the  equation  
 
applicable  to  the  transmitter  frequency.  
 
Recommended  separation  distance  
 
     
 
d=  1.2√P  80  MHz  to  800  MHz  
d=  2.3√P  800  MHz  to  2.7  GHz  
 
 
Where  P  is  the  maximum  output  power  rating  of  the  
transmitter  in  watts  (W)  according  to  the  transmitter  
 
manufacturer  and  d  is  the  recommended  separation  
 
distance  in  metres  (m).  
 
Field  strengths  from  fixed  RF  transmitters,  as  determined  
 
by  an  electromagnetic  site  survey,
the  compliance  level  in  each  frequency  range.
 
 
Interference  may  occur  in  the  vicinity  of  equipment  
marked  with  the  following  symbol:  
     
10  V/m  
 
 
27  V/m  
28  V/m  
 
9  V/m  
 
28  V/m  
 
28  V/m  
 
9  V/m  
and  the  e-­scope®  
(m)  
150  KHz  to  80  MHz  
0.12  
0.12  
0.38  
0.38  
1.2  
1.2  
3.8  
3.8  
12  
12  
Electromagnetic  environment  -­  Instructions  
a
 should  be  less  than  
80  MHz  to  800  MHz  
800  MHz  to  2.7  GHz  
0.23  
0.73  
2.3  
7.3  
23  
 
b
 
12  
0,23  
0,73  
2,3  
7,3  
23  
19