4.11
Device Configuration
Upon switching the device on, if at least
one handpiece is connected, the settings
selection screen will appear.
The MT family, together with the relevant
list of inserts, is the default setting. To
change insert family, press the relevant
insert family button;
Select the desired insert;
For each insert, the system automatically
set the optimal working settings. By using
the "-" and "+" buttons, the irrigation and
power levels can be changed within the
range defined by the manufacturer.
WARNING: To avoid overheating of
the treated parts, as the power increases,
it is recommended to proportionally
increase the irrigation level.
Fill the irrigation circuit before activating
the device, using the pump button. The
cycle can be interrupted as soon as the
liquid flows out of the insert, by pressing
the pump button again;
INSTALLATION AND CONFIGURATION
1
2
3
4
25