WARNING: Check the level of
physiological solution contained in the
physiological irrigation bag. Replace the
physiological irrigation bag with a new one
before it is completely empty.
WARNING: Sterile single-use material.
Before surgery. Before using any insert,
check the sterile package integrity. Insert
loses sterility in case of broken or damaged
packaging.
WARNING: Before surgery, assure
that the insert selected has been correctly
tightened onto the handpiece. The insert
is properly placed when PIEZOSURGERY®
medical torque wrench, used to tighten it,
produces a mechanical "CLICK" sound.
WARNING: The PIEZOSURGERY®
medical irrigation kit is guaranteed for a
single use, only. Separate and dispose of
this kit according to the current regulations
governing hospital waste.
WARNING: Before using the device,
check that the irrigation tube clamps are
open. After surgery, close the clamps before
disconnecting the irrigation kit from the
physiological irrigation bag.
CAUTION: Never force the cord-
handpiece connector into its port on the
device as the cord-handpiece connector
and/or the device can be damaged. If the
connector cannot be easily inserted into
the port, probably the connector and the
port do not correctly match. Ensure that the
connector index point of the cord-handpiece
is placed upwards.
CAUTION: Check that the
PIEZOSURGERY® medical handpiece is
correctly connected before using the system.
WARNING: Before each treatment, assure
that the insert suitable for the treatment has
been mounted on the handpiece. Only use
the PIEZOSURGERY® medical torque wrench
to tighten the insert onto the handpiece.
Do not use any other tools, such as pliers,
pincers, etc.
INSTALLATION AND CONFIGURATION
WARNING: Check that the insert
mounted into the handpiece is the same
one as shown on the display, otherwise,
performance will be affected.
CAUTION: To avoid damaging the foot
pedal cable, connect and/or disconnect it,
holding it by the connector only. Never pull
the cable.
CAUTION: Do not twist or rotate the
foot pedal cable connector when inserting
or removing it. Twisting can damage the
connector.
CAUTION: The foot pedal has been
specifically designed for the use with the
PIEZOSURGERY® plus device, only. Only use
the original foot pedal, otherwise, damages
or malfunctions may occur.
CAUTION: For a correct use of the device,
it is necessary to press the pedal and start
working with the insert not in contact with
the part to be treated, so that the electronic
circuit can recognize the best resonance
point of the insert without interference,
allowing its optimal performance.
WARNING: The patient must not come
into contact with the device chassis or the
foot pedal.
WARNING: Do not change the insert
when the handpiece is operating; the
operator may be injured.
WARNING: Be very careful to the sharp
edges of cutting inserts. When tightening and
removing these insert types, the edges may
cause injuries.
29