afriso DIT 10 Notice Technique page 47

Indicateur numérique de niveau de réservoir
Masquer les pouces Voir aussi pour DIT 10:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5.8
Mounting with cable gland (PG 9)
DAMAGE TO THE PRODUCT AND THE FACILITY
Only use existing tank connections to install the product.
Do not drill directly into the tank, but only into existing mounting flanges, caps
or blind connections.
• Verify that no foreign matter such as drilling chips can get into the tank during
mounting.
Failure to follow these instructions can result in equipment damage.
Mounting to a mounting flange with union nut, in a screw cap or in a free blind
connection.
A
Mounting
Cable gland (PG 9) with hex nut
NOTICE
1. Remove the mounting flange (B),
the cap or the blind connection
from the tank and drill a 15 mm
hole.
2. Insert the enclosed cable gland
(PG 9) (A) and fasten it with the
B
enclosed nut.
3. Route the cable of the pressure
sensor trough the cable gland (PG
9) (A) and tighten the cable gland
so that the cable can no longer be
moved and that the connection is
odour-tight.
EN
19
DIT 10

Publicité

Table des Matières
loading

Table des Matières