5.4.2
Inserting/connecting the battery
DAMAGE TO THE PRODUCT
•
Do not short-circuit the lithium battery.
• Verify correct polarity when plugging the battery connector into the socket on
the PCB in order to avoid a short circuit.
Failure to follow these instructions can result in equipment damage.
Mounting
NOTICE
After the pressure sensor and the
control unit have been electrically
connected, open the housing of the
control unit.
A
1. To do so, turn the upper part of the
housing vis-à-vis the lower part all
the way to the stop and then pull
B
the upper part of the housing off of
the lower part.
2. Press the mini pushbutton (A) and
hold it down.
3. Plug the two-pole battery connec-
tor into the two-pole socket (B) on
the PCB.
- Verify correct polarity when plug-
4. Release the mini pushbutton (A).
ging in the connector.
DIT 10
EN
14