Manufacturer's Declaration Tables

Guidance and manufacturer's declaration - electromagnetic emissions
The OCS2 is intended for use in the electromagnetic environment specified below. The customer
or the user of the OCS2 should assure that it is being used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-3
Voltage
fluctuations/flicker
emissions
IEC 610-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The OCS2 is intended for use in the electromagnetic environment specified below. The customer
or the user of the OCS2 should assure that it is being used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fat
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
or power supply
input lines
IEC 610-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c mains voltage prior to application of the test level.
T
Compliance
The OCS2 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
Group 1
likely to cause any interference in nearby electronic
equipment.
Class B
The OCS2 is suitable for use in all establishments other
NA
than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
NA
IEC 60601 test
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
<5 % U
(>95% dip
T
in U
) for 0.5 cycle
T
40 % U
(60% dip
T
in U
) for 5 cycles
T
70 % UT (30% dip
in U
) for 25 cycles
T
<5 % U
(>95% dip
T
in ) for 5 sec
3 A/m
3 A/m
Electromagnetic environment - guidance
Compliance
Electromagnetic environment -
level
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
NA
of a typical commercial or hospital
environment
Mains power quality should be that
NA
of a typical commercial or hospital
environment
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
OCS2 requires continued operation
NA
during power mains interruptions, it
is recommended that the OCS2 be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.
OSC2 Operator's Manual