Cleaning And Maintenance; Sterilization And Cleaning; Maintenance - Integra OCS2 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18

Cleaning and Maintenance

Sterilization and Cleaning

Ojemann Cortical Stimulator (OCS2)
Wipe the OCS2 clean with Clorox wipes or equivalent. Exercise care to avoid the ingress
of liquids into the enclosure of the device.
Ojemann Cortical Stimulator Probe (
The Ojemann Cortical Stimulator Probe (OCS2PND) is supplied non-sterile and must be
cleaned and sterilized before use.
Initial Probe Sterilization:
impregnated with isopropyl alcohol. Wrap and sterilize the product in the following
validated hospital sterilization cycle:
-
The OCS2PND can be steam sterilized by autoclave at 134ºC (273ºF) for 18
minutes.
-
The OCS2PND can also be sterilized at low temperature gas plasma. Sterilization
was validated with Sterrad 100S Sterilization System (Advanced Sterilization
Products). The validated cycle is: 10 min pre-exposure plasma phase followed
by two identical exposure phases (6 min injection, 2 min diffusion and 2 min
plasma).
Decontamination / Resterilization:
Probe with a non-aldehydic disinfectant. Decontaminate according to the following
procedures: Soak product 20 minutes in a non-aldehydic alkaline detergent. Rinse
with non-pyrogenic sterile water. Soak product one hour in Sodium Hydroxide 1N
solution. Rinse with non-pyrogenic sterile water. Autoclave 18 minutes at 134ºC
(273ºF).
Integra LifeSciences recommends that information on the original Lot Number be noted
on a label affixed to the package of the hospital-resterilized product.
Do not flash autoclave the Ojemann Cortical Stimulator Probe
CAUTION:
(OCS2PND). To prevent deformation, the OCS2PND probe should be
sterilized horizontally without constraint.

Maintenance

Ojemann Cortical Stimulator (OCS2)
The OCS2 contains no user serviceable parts. If maintenance is required, call Customer
Service and they shall provide the instructions to ship the unit to Integra LifeSciences
Corporation for repair.
Ojemann Cortical Stimulator Probe (OCS2PND)
The probe requires no maintenance as it is reusable up to a maximum of 5 uses and then
must be discarded. Before each use, inspect the device for evidence of connector
distortion or insulation deterioration such as cracks or fraying. The probe should not be
used if such deterioration is discovered.
8
OCS2 Operator's Manual
OCS2PND
Gently wipe the device with a lint free material
Thoroughly scrub the Ojemann Cortical Stimulator
)

Publicité

Table des Matières
loading

Table des Matières