Télécharger Imprimer la page

Chopin Aqua-TR Mode D'emploi page 58

Masquer les pouces Voir aussi pour Aqua-TR:

Publicité

Les langues disponibles

Les langues disponibles

GB
Aqua-TR
2.2 Removal of transport protection
■ The measurement cell is locked during transport with a
protection. It is recommended to keep the protection after removal
with the package and to use it for any transfer.
To remove the protection, proceed as follows :
Remove the drawer
Push down and remove it
2.3 Installation of transport protection
■ To install the protection, turn over the apparatus and verify the
traps of the cell are opened (if not, switch on the Aqua-TR and wait
for the end of the selfchecking sequence). Put the protection as
described below, then insert the drawer.
1. Insert the protection under the grain deflector
2. Enter the sides into the cell.
3. Block the emptying motor
User's manual
58
07/2015

Publicité

loading