WARNING
Incorrect device feature/settings!
Incorrect device features/settings may result in danger
to life and/or risk of explosion.
► Before every use, check whether the display
elements, the alarm features and information are
displayed correctly. If one of the items listed above
does not feature correctly or is incorrect, do not use
the gas detector and have it inspected.
The following features are active during the sensor
warm-up phase:
– The measured values flash
– The yellow LED is illuminated
– A warning notice is displayed
 The gas detector is ready to measure when the
measured values no longer flash and the yellow
LED is no longer illuminated. The warning notice
may continue to be displayed if warnings are
pending. For further information see: "Opening
information", page 38.
No alarms are issued during the warm-up phase!
4.4.3 Switching off the gas detector
1. Hold down ▲ and ▼ simultaneously until the
displayed countdown expires.
 The visual, acoustic and vibration alarms activate
briefly.
 The gas detector is switched off.
Or
1. Select
in measuring mode and confirm the
dialog.
2. Select and confirm Switch off.
The gas detector may only be switched off without a
prior sign-on if the Switch-off allowed feature is
activated using the Dräger CC-Vision PC software.
Factory setting: activated
The gas detector switches off automatically when it is
placed in the charging cradle (factory setting). This
setting can optionally be deactivated.
X-am
®
3500
If the gas detector remains off for longer than
21 days and is not charged, deep sleep mode is
activated. In deep sleep mode, the gas detector can no
longer be switched on using the Dräger CC-Vision PC
software or the Dräger X-dock. In this case, the gas
detector must be switched on manually.
4.5 Log user in or out
The gas detector has four configurable user levels. The
user levels can be set using the Dräger CC-Vision PC
software. User level 0 means that the user is not signed
on. User levels 1 to 3 require a password for sign-on.
The following passwords are created by default:
User level 1:
User level 2:
User level 3:
Dräger recommends changing the predefined
passwords after initial commissioning.
Default setting:
Function
Bump test
Fresh air adjustment
1)
Settings menu
1)
Maintenance menu
1) Maintenance and settings menu in user level 0 are not part of
the technical suitability test.
To log in a user:
1. Select
in measuring mode and confirm the
dialog.
2. Select and confirm Log in.
3. Enter the four-digit user password and confirm each
digit.
To sign out a user:
1. Select
in measuring mode and confirm the
dialog.
2. Select 'Sign out' and confirm the dialog.
Operation
0001
0002
0003
User level
0
-
-
-
-
-
-
-
en
-
-
-
-