This symbol indicates information that may
make the use of this product easier.
2.4 Glossary
Term
Description
Operation sig-
A periodic optical (green LED) and/or
nal
acoustic signal.
Monitoring
Monitoring without pump (diffusion)
Monitoring with pump (with pump
adapter)
Capture range
The capture range refers to a mea-
sured value range within which minor
variations in measured values (such
as signal noise, variations in concen-
tration) do not cause variations in the
display. Measured values outside the
capture range will be displayed with
the actual measured value.
Peak
Peak value
Quick bump
Test for alarm triggering.
test
Extended bump
Test for accuracy and alarm trigger-
test
ing.
Special state
If a special state is signalled, the user
is not warned of gas concentrations
which may be dangerous.
The following device features are
special states:
First set-up/configuration with the
PC, switch-on sequence, menu,
preparation phase of the wizards,
maintenance wizard bump test and
calibration, warm-up 1 of the sensors,
device error, measuring channel
error.
D-Light
The D-Light feature allows the user to
check and indicate compliance with
certain settings.
Physical sen-
CatEx-type sensors are designated
sors
as physical sensors. There are also
the electrochemical sensors.
X-am
®
3500
2.5 Abbreviations
Abbreviation Explanation
A1
Pre-alarm
A2
Main alarm
STEL
Short time exposure limit, threshold of
an exposition over a short period of
time (generally 15 minutes).
TWA
Time weighted average, average shift
values are generally limited to eight
hours exposure per day per workplace
for 5 days a week during a work life-
time. Observe the national definition of
the occupational exposure limit.
3
Description
3.1 Product overview
The graphics are displayed on the fold-out page.
Graphic A
1 Display
2 Locking screw for an
additional charging
module
3 Power supply unit
Graphic B
1 Gas inlets
2 Thread port for pumps
and calibration adapter
Graphic C
1 LED green/yellow/red
Graphic D
1 feature button 1
2 Feature button 2
Graphic E
1 Status information
Graphic F
1 Clip (optional)
Description
4 Charge LED green/red
5 Induction charger
3 Horn
4 Pump outlet and inlet
2 LED yellow/red
3 Feature button 3
2 Navigation bar
en

Publicité

loading