9.6 Tableau des alarmes
ALARME
E00
EEPROM ALARM
E01
BLACKOUT
E02
DOOR OPEN
E07
NO LOCKED DOOR
E09
NO UNLOCKED DOOR
E11
DIRTY WATER
E20
NO COLD WATER
E21
NO HOT WATER
E22
NO DEMI WATER
E26
PRINTER ERROR
E30
FLOWMETER 1
E31
FLOWMETER 2
E32
FLOWMETER 3
E34
PRODUCT TIMEOUT
E41
DRAIN TIMEOUT
E43
DRY TEMPERATURE
DESCRIPTION
Firmware installed successfully.
A blackout occurred during cycle
execution and the program was stopped.
The door is open or unlocked.
Close the door.
The door did not lock within the set
time.
The door did not unlock within the set
time.
Check for dirt on the load and check the
inlet demineralized water.
The cold water tap could be closed or
partially open. Check it.
The hot water tap could be closed or
partially open. Check it.
The demineralized water tap could be
closed or partially open. Check it.
The printer is turned off, unplugged or
out of paper. Check it.
Chemical liquid tank 1 might be empty.
Check it.
Chemical liquid tank 2 might be empty.
Check it.
Chemical liquid tank 3 might be empty.
Check it.
The density of the chemical liquid might
be too high. Check it.
The drain pipe might be blocked or
clogged. Check it.
The load might not be dried because the
minimum set temperature has not been
reached.
ACTION
1-Appuyer sur RÉINITIALISER
2-Installer nouveau jeu de
données
Appuyer sur RESET
Appuyer sur RESET
Si elle persiste appeler le tech-
nicien.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Check if the safety floatswitch
has not tripped.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
Press RESET
If it persists, call the technician.
49