Power Frequency
Magnetic Fields
30 A/m
(50/60Hz) IEC 61000-
4-8
NOTE : Uт is the main voltage (AC) prior to the application of the test level.
Guidance 2
Recommended Separation Distances Between Portable and Mobile Communication Equipment and the i700 wireless
Rated Maximum
Output Power of
the Transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the ransmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 : At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Guidance 3
The i700 wireless system is intended for use in the electromagnetic environment specified below. The customer or the user of the i700 wireless
system should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
3 Vrms 150 kHz to 80 MHz
Outside ISM Bands amateur
Conducted RF
IEC 61000-4-6
6 Vrms 150 kHz to 80 MHz
In ISM Bands amateur
Radiated RF IEC
3 V/m 80 MHz to 2.7 GHz
61000-4-3
NOTE 1 : At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
NOTE 3 : The ISM (Industrial, Scientific, and Medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz;
13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
30 A/m
Separation Distance According to the Frequency of Transmitter [M]
150 kHz to 80 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Compliance Level
3 Vrms
6 Vrms
3 V/m
Power frequency magnetic fields
should be at levels characteristic of a
location in a typical commercial or
hospital environment.
IEC 60601-1-2: 2014
80 MHz to 2.7 GHz
d = 2.0 √P
Electromagnetic Environment – Guidance
Portable and mobile RF communications
equipment should not be used closer to any part
of the Ultrasound System, including cables, than
the recommended separation distance. This is
calculated using the equation applicable to the
frequency of the transmitter.
Recommended Separation Distance (d):
d = 1.2 √P
IEC 60601-1-2:2007
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 80 MHz to 2.5 GHz
IEC 60601-1-2:2014
d = 2.0 √P 80 MHz to 2.7 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer, d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol :
Intraoral Scanner i700 wireless
0.20
0.63
2.0
6.3
20
15