SETTING THE DATE AND TIME

( PLEASE PERFORM THE FOLLOWING STEPS WITH DOOR OPENED )
To set the date and time which are shown on the display, proceed as follows:
1) Turn on the autoclave with the "ON/OFF" main switch, located on the right side of the control panel and
keep pressing the "START/STOP" button. After some seconds the "TIME/MESS." display shows the writing -
1-. This means you accessed the first programming page and you can now release the "START/STOP"
button.
The "TEMPERATURE" display shows the blinking seconds. Press the "14A" button, to access selection of
the minutes (MM display). To change the values displayed on "MM" press the "13A" button or the "16A"
button to increase or respectively decrease them.
Now, press the "14A" button, to enter hours selection (H display). The minutes stop blinking and the hours
start blinking. Set the hours to the desired time data pressing the "13A" button or the "16A" button, to
increase or respectively decrease the current setting.
Press "14A" to enter setting of the day (DATE display). Press "13A" or "16A" to increase or respectively
decrease the current day value. Press "14A" to enter setting of the month (DATE display). Press "13A" or
"16A" to increase or respectively decrease the current month value.
Press "14A" to enter setting of the year (DATE display). Press "13A" or "16A" to increase or respectively
decrease the current year value.
2) After setting the new date and time, press the "START/STOP" button, to confirm the data and enter the
next programming page. The "TIME/MESS." display shows the writing -2-. On this page, it is possible to align
the atmospheric pressure. The "PRESSURE" display shows the blinking message 0.
Press the
"START/STOP" button, to confirm the new value. This setting is required to align the atmospheric pressure
which changes based on the local altitude. An acoustic alarm warns you when the new data are saved.
3) Now, the system automatically enter the next programming page -3-. Here it is possible to set the type of
external units connected to the system (TEMPERATURE display). Pressing "13A" button or the "16A" button
to increase or respectively decrease the number of connected units: 0=None – 1=Internal Printer –
2=External Printer – 3=Computer – 4=Internal Printer+Computer.
4) Press the "START/STOP" button to confirm the new data and enter the following programming page –4- (
"TIME/MESS." display.). This page shows the autoclave serial number. Do not change this setting. Only
authorised TECNO-GAZ staff may change this setting. Turn off the "ON/OFF" main switch, to exit this
menu.