Specifications - Philips RESPIRONICS SimplyGo Mini Mode D'emploi

Masquer les pouces Voir aussi pour RESPIRONICS SimplyGo Mini:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

SimplyGo Mini User Manual

Specifications

Operating
Conditions
Storage and Transport
Conditions
Expected Service Life
of the device and
accessories
Oxygen
Concentration
Flow Settings and
Pulse Volumes
52
Operational temperature:
41° F to 95° F (5° C to 35° C)
Relative humidity: 15% to 93%
Atmospheric Pressure: 70 kPa to 101 kPa
Altitude: 0 to 10,000 ft� (3048 m�)
-4° F to 140° F (-20° C to 60° C) - unit only
Relative humidity: up to 93%, non condensing
5 years
90% +6% - 3% at all settings
Per ISO 80601-2-67 test method
+/-15% at STPD*
+/-25% over the rated environmental range
*STPD is 101�3 kPa at an operating temperature of 20 °C, dry
Average of 30 consecutive pulses with 1�5 m length tubing