Recommended Separation Distances; Table 15: List Of Recommended Separation Distances - Nikon Optos California P200DTx rg Manuel D'utilisation

Ophtalmoscope panoramique
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Chapter 4 - Technical Specification
IEC60601 compliance test
Immunity test
level
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment in the location due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
device is used exceeds the applicable RF compliance level above, the device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the device.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.

4.3.3 Recommended separation distances

This section lists the recommended separation distances between portable and mobile communication
equipment and the device.
The
P200DTx
is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled.
The customer or the user of the
minimum distance between portable and mobile RF communication equipment (transmitters) and the
as recommended below, according to the maximum output power of the communications equipment.

TABLE 15: List of recommended separation distances

Rated maximum output power
of transmitter
(W)
00.1
0.1
1.0
10
100
For transmitter rated at a maximum output power
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power of the transmitter in watts (W) according to the transmitter manufacturer.
Note
At 80MHz and 800MHz the higher frequency range applies.
l
These guidelines may
l
absorption and reflection from structures, objects and people.
Page 62 of 69
English
Electromagnetic environment-guidance
P200DTx
can help prevent electromagnetic interference by maintaining a
Separation distance according to frequency of transmitter
150kHz to 80MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
listed above, the recommended separation distance d in
NOT
apply in all situations. Electromagnetic propagation is affected by
NOT
80MHz to 800MHz
800MHz to 2.5GHz
d=1.2√P
d=2.3√P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
Part Number: G108025/4GCA
Copyright 2017, Optos plc. All rights reserved.
P200DTx

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Optos california p200dtx faOptos california p200dtx icg

Table des Matières