1. INTRODUCTION

This manual was created to help you install and learn all about the functional possibilities of the BCSD scale.
The scales should be sited in a location that will not degrade the accuracy.
Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning vents.
Avoid unsuitable tables. The tables or floor must be rigid and not vibrate. Do not place near vibrating machinery.
Avoid unstable power sources. Do not use near large users of electricity such as welding equipment or large motors.
Avoid high humidity that might cause condensation. Avoid direct contact with water. Do not spray or immerse the scales in
water.
Avoid air movement such as from fans or opening doors. Do not place near open windows.
Keep the scales clean.
Do not stack material on the scales when they are not in use.

2. MAIN TECHNICAL SPECIFICATIONS

Model
BCSD6
BCSD15
BCSD30
Stabilisation Time
Operative Temperature:
Power supply:
Battery life:
Recharge time:
Parameter Setup:
Display :
Case:
Programmable Zero range:
Programmable Auto zero at start-up:

2.1 SPECIFICATIONS FOR THE REMOTE SCALE

Load cell drive voltage:
Range of input signal
Zeroing field:
Minimum voltage per division:
Internal resolution:
Load cells:
Connection of load cells:
Maximum cable length:
Programmable Zero range:
Automatic zero at start-up:
Plate
Capacity
dimensions
mm
230 x 300
230 x 300
230 x 300
2 Seconds typical.
0°C - 40°C / 32°F - 104°F.
with fitted 240 Vac mains power adapter or with fitted battery
up to 70 hours of continuous use.
12 hours.
Calibration fully digital and programmable from the keyboard.
3 LCD backlit with 6 digits
ABS Plastic.
+/- 0, 2, 4, 10, 50 or 100 % of max load capacity.
+/- 0, 2, 5, 10 or 20% of max load capacity.
Max 5V/150mA.
0 mV/V - 3 mV/V.
0 - 5 mV.
0,02 V
600.000 points
up to 4 load cells with input resistance of 350 Ohm
9-pin connector put on the scale.
6 meters
+/- 0, 2, 4, 10, 50 or 100 % of max load capacity.
+/- 2% of the max. capacity
WARNINGS
Standard
Division
kg
g
6
0,1
15
0,2
30
0,5
.
4
Internal sensitivity
recommended
g
0,02
0,05
0,1
Min. APW
g
0,1
0,2
0,5