Troubleshooting; Alarm Troubleshooting - ResMed Stellar Serie Guide De L'utilisateur

Masquer les pouces Voir aussi pour Stellar Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact ResMed.

Alarm troubleshooting

The most common reason for an alarm to sound is because the system has not been properly assembled.
Check that the air tubing has been properly attached to the device and patient interface (and humidifier if
used).
Notes:
The alarm log and alarm settings are maintained when the device is powered down and in the event of
a power loss.
If multiple alarms are active simultaneously, the alarm with the highest priority will be displayed first.
If an alarm activates repeatedly, discontinue use and return the device for servicing.
Problem / possible cause
LCD: Internal Battery Empty!
The remaining battery charge is
below 15%. The device can be
powered by the internal battery
for maximum 2 minutes.
LCD: System Failure!
Component failure.
The device stops delivering air
pressure (system failure 6, 7, 9,
22, 38). Therapy cannot be
started (system failure 21).
The device internal temperature
is too low for the device to
start/deliver therapy (system
failure 21).
The self test of the device fails
and therapy cannot be started
(system failure 21).
Component failure (system
failure 8, 25)
LCD: Over Pressure!
The device generates a pressure
that is greater than 59 cm H
Treatment will be stopped.
LCD: Blocked Tube!
Air path is blocked.
Action
Connect the device to mains power.
Note: In case of a total power failure, the therapy settings will be
stored and therapy will resume when the device is powered again.
1. Power-off the device.
2. Power-on the device again.
1. Ensure that ambient temperature is above 5°C.
If the device has been stored below 5°C, allow sufficient time for it
to acclimatise before use.
2. Power-off the device.
3. Power-on the device again.
If the problem persists, return the device for servicing.
1. Power-off the device.
2. Power-on the device again.
If the problem persists, return the device for servicing.
1. Power-off the device.
2. Power-on the device again.
1. Power-off the device.
O.
2. Check that the air tubing is connected properly.
2
3. Power-on the device again.
4. Start Learn Circuit function.
Note: If the alarm activates repeatedly internal components may be
defective. Discontinue use and return the device for servicing.
1. Check the air path for any blockages.
2. Remove the blockages.
3. If the alarm is not cleared, stop treatment.
4. Re-start treatment.
English
25

Publicité

Table des Matières
loading

Table des Matières