Info menu

Event Summary

Used Hours

Device Information

Reminders

Event summary displays the summary of three types of events:
changes in settings, alarms and system events (eg, connection of
ResMed USB stick). There are up to 200 events of each type,
displayed in chronological order, with the most recent event
displayed at the top by default.
Note: When the capacity of 200 logged events is reached, the oldest
log record is overwritten.
Used Hours during the last seven days of treatment are displayed in
a bar graph and can be compared to the data of the last 365 days.
This screen shows the serial number (displayed as Device SN),
device number, software version and other component versions.
Data on this screen may be requested for servicing or as part of
problem-solving by a technician.
The Clinician uses the Reminders menu to alert you to specific
events; for example, when to replace the mask, when to replace the
filter and so on. The reminder appears in yellow as the date
approaches (within 10% of the reminder period). The reminder also
displays when the device is powered on. You can clear a reminder
message by selecting 'Reset', which clears the current reminder
date to OFF, or displays the next pre-set Reminder date.