ROUBLESHOOTING
If there is a problem, try the following suggestions. If the problem cannot be solved,
contact your equipment supplier or ResMed. Do not attempt to open the unit.
Problem
No display.
Insufficient air delivered
from AutoSet Spirit.
AutoSet Spirit does not
start when you breathe
into the mask.
Possible Cause
Power not connected or switch at
back is not on.
Ramp Time or Settling Time is in
use.
Air filter is dirty.
Air tubing is kinked or punctured.
Air tubing not connected
properly.
Mask and headgear not
positioned correctly.
Plugs missing from access port on
mask.
Pressure required for treatment
may have changed.
Power cord not connected
properly.
Power outlet may be faulty.
AutoSet Spirit unit not switched
on.
SmartStart not on.
Leak Alert has been enabled;
SmartStart has automatically been
disabled.
Use of a Mirage Full Face Mask.
Breath is not deep enough to
trigger SmartStart.
Solution
Ensure the power cable is
connected and that the switch at
the back of the unit is in the ON
position.
Wait for air pressure to build up.
Replace air filter.
Straighten or replace tubing.
Check air tubing.
Adjust position of mask and
headgear.
Replace plugs.
See your clinician to adjust the
pressure.
Connect power cord firmly at both
ends.
Try another power outlet.
Switch power switch at rear of
AutoSet Spirit to ON.
Enable SmartStart.
Disable Leak Alert to enable
SmartStart.
SmartStart does not work with a
full face mask as the anti-asphyxia
valve will not allow sufficiently
high pressure on exhalation.
Take a deep breath in and out
through the mask.
T
ROUBLESHOOTING
29