Operating the field indicator
!
48
To configure the device, first remove the housing cover. Then remove the display from the
electronics unit. The keys (+, -, E) can now be accessed.
During configuration, the display must be connected to the electronics unit. Afterwards, the cable
can be removed and the display positioned at the desired angle.
6.2.1
Navigation
The operating fields are split into 2 levels.
Menu: Various menu items can be selected on the menu level. The individual menu items are an
aggregation of the associated operating functions.
Operating function: An operating function can be regarded as an aggregation of the operating
parameters. The operating functions are used to operate and configure the device.
Operating keys:
'E' Enter key: Press and hold down the E key for longer than 3 seconds to enter the programming
menu.
– Selecting operating functions.
– Accepting values.
– If the E key is pressed for longer than 3 seconds, the system returns directly to the Home position.
Beforehand, you are asked whether the data entered up to now should be saved.
– Saving data entered.
Selection keys '+/-':
– Selecting the menus.
– Configuring parameters and numerical values.
– After selecting the operating function, the value is entered or the setting changed by pressing the
+ or - keys.
!
Note!
If the keys are pressed for an extended period, the digits are changed with increasing speed.
Note!
If the + or - keys are pressed in the "Program Name" and "Program Version" operating position, the
display is scrolled horizontally as these positions (7-digit) cannot be displayed completely in the 14-
segment display.