To Troubleshoot The Frx Trainer - Philips HeartStart FRx Trainer Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

16
behavior
Trainer says
ANALYZING before
pads are applied to
manikin.
Trainer does not say
ANALYZING when pads
are applied to the
manikin adapter.
Trainer's voice
instructions are
distorted.
Battery depletes
prematurely.
FRx TRAINER 861306

To troubleshoot the FRx Trainer

The following table provides troubleshooting tips.
possible cause
• User touched the adhesive gel
on the pads and activated the
sensing circuitry of the Trainer.
• Pad wires are twisted.
• Pad wires are not separated.
• Pads are touching each other.
• Pads placement is incorrect.
• Pads are old, damaged, or use
discontinued blue gel.
• Manikin adapter is broken.
• In scenarios 3 and 8, pad has not
been removed and repositioned.
• Pad wires are twisted.
• One of the batteries is installed
backwards.
• Trainer is stored so that On/Off
button is inadvertently pressed.
• One of the batteries is installed
backwards.
recommended action
• Avoid touching the adhesive gel
on the pads.
• Untwist the pad wires.
• Separate the pad wires between
the pads and the red label.
• Separate pads.
• Check for proper pad and
manikin adapter placement.
Remove and reposition pads or
adapter if indicated.
• Replace training pads.
• Replace manikin adapter.
• Remove and reapply one pad.
• Untwist the pad wires.
• Reinstall batteries according to
the diagram in the battery
holder.
• Store the Trainer so as to
prevent inadvertent pressure on
the On/Off button.
• Reinstall batteries according to
the diagram in the battery
holder.