SERIAL Operation

The SERIAL is set in order to operate by connecting the controller to an external
NOTE
PC that uses the "HYPER TERMINAL" program of Windows. To operate in serial
mode, connect the serial port to the controller "Serial" connector on the rear
panel, run the Windows "HYPER TERMINAL" program on the PC with the
following settings:
This way of operation will allow to verify the functionality of the unit in the
"SERIAL" operating mode, give the commands and get all the operating data.
See the "Technical Information" paragraph for details.

START/PROTECT Operation

The controller can operate in two modes: Start and Protect.
In Start, the 3 output channels can provide all the output power
regardless of the pump condition up to the short circuit condition.
This operating mode must be used to start the pump at high pressure.
In Protect, the controller limits the output current and switches off
the HV when the current exceeds the threshold value (8 mA) for more
than 2 seconds.
Ion Pumps Control Unit (No Battery) User Manual / 87-900-116-01 (E)
COM1
9600 baud
No parity
8 data bits
1 stop bit
Installation Procedure
Operating Modes